Olympus

Objective lens Group for Olympus GIF-H260 gastroscope

$0.00

► Brief Description

The objective lens system in the Olympus GIF-H260 gastroscope is a sophisticated optical component located at the very tip of the endoscope. It works in conjunction with the CCD (Charge-Coupled Device) sensor to capture high-quality images of the upper digestive tract.

► Features

  • Superior Image Quality
  • Extra-Wide Field of View
  • Broad Depth of Field
  • Spectral Integrity
  • Clear View Maintenance

► Specification

Brand Olympus
Model GIF-H260
Series Objective lens
Field of View 140°
Depth of Field 3 – 100 mm
Distal End Diameter 9.8 mm
MOQ 1 Piece
Package Carton / Standard Export Packing
Transportation DHL, FedEx, UPS, etc.
In Stock Yes
Price Accept Negotiation

Olympus A70940A Autoclavable Arthroscope for repair

$0.00

► Brief Description

Fault: 

1. Image blur, black dots and dark.

2. Fiber not light enough.

3. Rod lens broken.

4. Objective lens broken and dirty.

Solution: 

1. Replace rod lens.

2. Replace inner tube.

3. Replace outer tube.

4. Replace objective lens group.

5. Clean the eyepiece.

6. Replace fiber.

7. Clean and install the optical path, check the image.

8. Seal with endoscope and test.

â–ºProblem & Solutoin

Brand Olympus
Model A70940A
Fault 1. Image blur, black dots and dark. 2. Fiber not light enough. 3. Rod lens broken. 4. Objective lens broken and dirty.
Solution 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Clean and install the optical path, check the image. 8. Seal with endoscope and test.

► Common issues with A70940A Arthroscope

  • No image/Blackout
  • Dim image/Insufficient light
  • Fuzzy/Blurry image
  • Deformed image/Shading
  • Dust or spots in the image
  • Damaged insertion tube
  • Fluid ingress

Olympus BF-XP160F Bronchoscope for repair

$0.00

► Brief Description

Fault:

1. Image degumming, serious color, CCD aging, a few black and gray spots;

2. 20% of the light bundle is broken;

3. Insert tube aging slightly;

4. Biopsy tube wear, not smooth;

5. Bending section, steel wire, screw tube wear;

Solution:

Replace CCD, image beam, beam, insert tube, pliers tube, snake bone, steel wire, spiral tube, rubber.

â–ºProblem & Solutoin

Brand Olympus
Model BF-XP160F
Fault 1. Image degumming, serious color, CCD aging, a few black and gray spots; 2. 20% of the light bundle is broken; 3. Insert tube aging slightly; 4. Biopsy tube wear, not smooth; 5. Bending section, steel wire, screw tube wear;
Solution Replace CCD, image beam, beam, insert tube, pliers tube, snake bone, steel wire, spiral tube, rubber.

► Common issues with BF-XP160F Bronchoscope

  • Bending Section Issues
  • Insertion Tube Damage
  • Optical/Image System Problems
  • Fluid Leaks
  • Channel Obstructions
  • Light Guide and Illumination Issues

Olympus CCD For CF-H260AI Colonoscope

$0.00

► Brief Description

The Olympus CF-H260AI is a high-definition (HDTV) colonoscope from the EVIS EXERA II series. The CCD (Charge-Coupled Device) is the “eye” of the endoscope, located at the distal tip. It is responsible for converting light into electrical signals to produce the video image.

► Features

  • HDTV Image Capture
  • Narrow Band Imaging (NBI) Support
  • Light-to-Signal Conversion
  • Color Reproduction and Signal Processing

► Specification

Brand Olympus
Model CF-H260AI
Condition Working in Good Condition
Series CCD
Sensor Type High-Definition HDTV CCD
Image Resolution HDTV
Field of View (FOV) 140° (Wide-angle)
Depth of Field 3 mm to 100 mm
Aspect Ratio 4:3 or 16:9
MOQ 1 Piece
Package Carton / Standard Export Packing
Transportation DHL, FedEx, UPS, etc.
In Stock Yes
Price Accept Negotiation

Olympus CF-240AI Colonoscope for repair

$0.00

► Brief Description

Fault: 

Image blur,insert tube aging,light guide squashed,light guide broken 50%,nozzle blocked,angle tight,forceps channel aging.

Solution:

Replace insertion tube,CCD chips,light guide,forceps channel,water channel,angulation wire,bending mesh,spiral tube,light guide tube,bending rubber.

â–ºProblem & Solutoin

Brand Olympus
Model CF-240AI
Fault Image blur,insert tube aging,light guide squashed,light guide broken 50%,nozzle blocked,angle tight,forceps channel aging.
Solution Replace insertion tube,CCD chips,light guide,forceps channel,water channel,angulation wire,bending mesh,spiral tube,light guide tube,bending rubber.

► Common issues with CF-240AI Colonoscope

  • Insertion Tube Damage
  • Image Quality Issues
  • Mechanism Failures
  • Connector and Light Source Issues

Olympus CF-240L Colonoscope for repair

$0.00

► Brief Description

Fault: Leakage, glass of light bundle blur, rubber leak, angle not enough,image fuzzle,objective lens scratch.

Solution: Replace rubber, objective lens, C cover. Adjust the angle.

â–ºProblem & Solutoin

Brand Olympus
Model CF-240L
Fault Leakage, glass of light bundle blur, rubber leak, angle not enough,
image fuzzle,objective lens scratch,
Solution Replace rubber, objective lens, C cover. Adjust the angle.

► Common issues with CF-240L Colonoscope

  • Damaged Insertion Tube
  • Worn or Damaged Bending Section
  • Fuzzy or Blocked Image
  • Dark Spots or Broken Pixels
  • Poor Light Transmission
  • Fluid Leaks
  • Blocked Air/Water or Suction Channels
  • Connector Damage

Olympus CF-2T160L Colonoscope for repair

$0.00

► Brief Description

Fault: 

1. With image.

2. 10% light bundle broken.

3. Insert tube is ok.

4. Biopsy tube abrasion.

5. Angle tied, bending section transformation.

6. Wires, helix tube abrasion, no up angle.

Solution:

Replace image bundle; insert tube; biopsy tube; bending mesh and wires; coil pipe; rubber

â–ºProblem & Solutoin

Brand Olympus
Model CF-2T160L
Fault 1. With image. 2. 10% light bundle broken. 3. Insert tube is ok. 4. Biopsy tube abrasion. 5. Angle tied, bending section transformation. 6. Wires, helix tube abrasion, no up angle.
Solution Replace wires, helix tube, bending section and rubber.

► Common issues with CF-2T160L Colonoscope

  • Bending Section Issues
  • Insertion Tube Damage
  • Optical/Image System Problems
  • Fluid Leaks
  • Channel Obstructions
  • Light Guide and Fiber Optic Issues

Olympus CF-H190L Colonoscope for repair

$0.00

► Brief Description

Fault:

1. Image ok;

2. Light bundle still ok;

3. The front end of the insert tube is broken, and the serpentine; the position of stress boot is leaking;

4. Biopsy tube wear;

5. Left Angle failure, bending section, wire and helix tube wear;

6. Soft and hard adjustment wire wear, relatively tight;

7. Rubber leakage.

Solution:

Replace the insertion tube, biopsy tube, bending section, wire, helix tube, soft and hard adjusting wire and rubber. (No warranty for CCD).

â–ºProblem & Solutoin

Brand Olympus
Model CF-H190L
Fault 1. Image ok; 2. Light bundle still ok; 3. The front end of the insert tube is broken, and the serpentine; the position of stress boot is leaking; 4. Biopsy tube wear; 5. Left Angle failure, bending section, wire and helix tube wear; 6. Soft and hard adjustment wire wear, relatively tight; 7. Rubber leakage.
Solution Replace the insertion tube, biopsy tube, bending section, wire, helix tube, soft and hard adjusting wire and rubber. (No warranty for CCD)

► Common issues with CF-H190L Colonoscope

  • No Image/Black Screen
  • Dim or Uneven Illumination
  • Blurry or Out-of-Focus Image
  • Stiff or Difficult Angulation
  • Leaking or Failed Leak Test
  • Abnormal Angulation
  • Biopsy Channel Obstruction
  • Damaged Bending Rubber
  • Buttons Not Responding
  • Water or Air Channel Issues

Olympus CF-Q160AL Colonoscope for repair

$0.00

► Brief Description

Fault:

1.No image, CCD broken;

2.image broken completely;

3.insert tube aging;

4.biopsy tube abrasion;

5.bending section abrasion;

6.wires and helix tube abrasion;

7.light guide tube aging and wrinkle.

Solution:

Replace CCD, light bundle, insert tube,biopsy tube,aire&water tube,bending section,wires,helix tube,light guide tube,all the fstener part,clamp and rubber.

â–ºProblem & Solutoin

Brand Olympus
Model
CF-Q160AL
Fault 1. With image. 2. 10% light bundle broken. 3. Insert tube is ok. 4. Biopsy tube abrasion. 5. Angle tied, bending section transformation. 6. Wires, helix tube abrasion, no up angle.
Solution Replace wires, helix tube, bending section and rubber.

► Common issues with CF-Q160AL Colonoscope

  • Bending Section Issues
  • Insertion Tube Damage
  • Optical/Image System Problems
  • Fluid Leaks
  • Channel Obstructions
  • Light Guide and Fiber Optic Issues

Olympus CHF-P20Q endoscope for repair

$0.00

► Brief Description

Fault:

1.Insert tube cracked;

2.Rubber leakage;

3.Image with black point and cracked;

4.Angulation tight;

5.Biopsy tube aging

Solution:

Replace image bundle; insert tube; biopsy tube; bending mesh and wires; coil pipe; rubber

â–ºProblem & Solutoin

Brand Olympus
Model CHF-P20Q
Fault 1.Insert tube cracked; 2.Rubber leakage; 3.Image with black point and cracked; 4.Angulation tight; 5.Biopsy tube aging
Solution Replace image bundle; insert tube; biopsy tube; bending mesh and wires; coil pipe; rubber

► Common issues with CHF-P20Q endoscope

  • Damaged or Punctured Insertion Tube
  • Issues with the Bending Section
  • Image Quality Problems
  • Fluid Leakage
  • Blocked Channels

Olympus CYF V2 Cystoscope For Repair

$0.00

► Brief Description

Fault: Leakage.

Solution:

Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber.

â–ºProblem & Solutoin

Brand Olympus
Model CYF V2
Fault Leakage
Solution Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber

► Common issues with CYF V2 Cystoscope

  • No Image / Dark Image
  • Poor Image Resolution / Blurry Image
  • Ineffective Suction
  • Difficulty with Angulation (Stiff or Loose)
  • Bubbles in the Field of View
  • Damaged Insertion Tube or Sheath

Olympus ENF-GP fiber nasopharyngeal laryngoscope for repair

$0.00

► Brief Description

Fault:

1. Image with black point and image bundle broken

2. Light bundle aged

3. Insert tube aging and wrinkle

4. Bending section deformation

5. Angle with not enough

6. Rubber abrasion

Solution:

Replace image bundle; light bundle; insert tube; wires; helix tube; bending section; rubber

â–ºProblem & Solutoin

Brand Olympus
Model ENF-GP
Fault 1. Image with black point and image bundle broken 2. Light bundle aged 3. Insert tube aging and wrinkle 4. Bending section deformation 5. Angle with not enough 6. Rubber abrasion
Solution Replace image bundle; light bundle; insert tube; wires; helix tube; bending section; rubber

► Common issues with ENF-GP Endoscope

  • Bending Section Issues
  • Insertion Tube Damage
  • Optical/Image System Problems
  • Fluid Leaks
  • Light Guide and Illumination Issues

Olympus GIF-1TQ160 gastroscope for repair

$0.00

► Brief Description

Fault:

1.With image, but CCD with broken.

2.Light bundle broken completely.

3.Insert tube abrasion.

4. No biopsy tube.

5. No air&water tube, no water supply tube.

6. No bending section assembly.

7. No digital tip.

Solution:

Replace CCD, digital tip, light bundle, insert tube,biopsy tube, air&water tube, water supply tube,bending section,wires,helix tube and rubber.

â–ºProblem & Solutoin

Brand Olympus
Model GIF-1TQ160
Fault 1.With image, but CCD with broken. 2.Light bundle broken completely. 3.Insert tube abrasion. 4. No biopsy tube. 5. No air&water tube, no water supply tube. 6. No bending section assembly. 7. No digital tip.
Solution Replace CCD, digital tip, light bundle, insert tube,biopsy tube, air&water tube, water supply tube,bending section,wires,helix tube and rubber.

► Common issues with GIF-1TQ160 gastroscope

  • Bending Section and Angulation Issues
  • Insertion Tube Damage
  • Biopsy/Suction Channel Blockage
  • Image Quality Problems
  • Air/Water Insufflation Issues
  • Fluid Leakage

Olympus GIF-H170 gastroscope for repair

$0.00

► Brief Description

Fault:

1. The top left and middle right side with blur image, objective Lens broken.

2. 20% light bundle broken, the small lens turned yellow.

3. Insert tube yellow and wrinkles.

4. The biopsy tube abrasion.

5. The wire and helix tube abrasion.

6. Light guide tube aging and wrinkle.

Solution:

Replace objective lens, light bundle, small lens, insert tube, biopsy tube, wires, helix tube, light guide tube and rubber.

â–ºProblem & Solutoin

Brand Olympus
Model GIF-H170
Fault 1. The top left and middle right side with blur image, objective Lens broken. 2. 20% light bundle broken, the small lens turned yellow. 3. Insert tube yellow and wrinkles. 4. The biopsy tube abrasion. 5. The wire and helix tube abrasion. 6. Light guide tube aging and wrinkle.
Solution Replace objective lens, light bundle, small lens, insert tube, biopsy tube, wires, helix tube, light guide tube and rubber.

► Common issues with GIF-H170 gastroscope

  • Insertion Tube Damage
  • Light Guide Fiber Breakage
  • Image Sensor Failure (CCD/CMOS)
  • Angulation Control Issues
  • Water/Air Channel Blockage
  • Biopsy Channel Blockage or Leakage
  • Distal Tip Damage
  • Fluid Leakage

Olympus GIF-H180J gastroscope for repair

$0.00

► Brief Description

Fault:

1.image with shadow;

2.light bundle broken;

3.insert tube with bite mark;

4.biopsy tube abrasion;

5.bending section serious deformation;

6.wires and helix tube abrasion;

7.light guide tube aging and wrinkle.

Solution:

Replace CCD, light bundle, insert tube,biopsy tube,aire&water tube,bending section,wires,helix tube,light guide tube and rubber.

â–ºProblem & Solutoin

Brand Olympus
Model GIF-H180J
Fault 1.image with shadow; 2.light bundle broken; 3.insert tube with bite mark; 4.biopsy tube abrasion; 5.bending section serious deformation; 6.wires and helix tube abrasion; 7.light guide tube aging and wrinkle.
Solution Replace CCD, light bundle, insert tube,biopsy tube,aire&water tube,bending section,wires,helix tube,light guide tube and rubber.

► Common issues with GIF-H180J gastroscope

  • Image and Optical Problems
  • Bending Section Issues
  • Air/Water Channel Problems
  • Suction Channel Problems
  • External Damage and Wear

Olympus GIF-Q150 Gastroscope for repair

$0.00

► Brief Description

Fault: 

1. With image;

2. Light bundlebreak badlyï¼›

3. Insert tubeaging;

4. Biopsy tubeabrasion, leakage badly;

5. Angle tied, bendingsection, wires and helix tube abrasion;

6. Rubberwithout glue.

Solution:

Replace light bundle, insert tube, biopsy tube, bending section, wires, helixtube and rubber. (No warranty for CCD)

â–ºProblem & Solutoin

Brand Olympus
Model GIF-Q150
Fault 1. With image; 2. Light bundlebreak badlyï¼› 3. Insert tubeaging; 4. Biopsy tubeabrasion, leakage badly; 5. Angle tied, bendingsection, wires and helix tube abrasion; 6. Rubberwithout glue.
Solution Replace light bundle, insert tube, biopsy tube, bending section, wires, helixtube and rubber. (No warranty for CCD)

► Common issues with GIF-Q150 Gastroscope

  • No image/Blackout
  • Frozen image
  • Poor image quality/Fuzzy image
  • Water leakage/Fluid ingress
  • Insufficient suction
  • Weak air/water flow
  • Bending section doesn’t move/Stiff bending section
  • Control knobs are loose/No response from control knobs
  • Damaged insertion tube
  • Damaged light guide
  • Stuck elevator mechanism
  • Dirty lens
  • Error message on processor

Fix Once Right, Save Replacement Costs

Flexible endoscopes are among the most advanced and most used devices in the healthcare environment. As their use continues to expand, they are a critical component in improving patient surgical outcomes.

However, without the right quality assurance services, they can also present infection risks to customers, leading to significant avoidable service costs over the life of the device.

Long-term replacement or good accessory maintenance is necessary to extend the life of your endoscope equipment, and CareU Tech is committed to reducing service costs and minimizing patient risk so that repaired endoscopes always meet optimal performance standards.

If you have any needs for flexible endoscope accessories, such as insertion tubes, light guides, biopsy channels, air/water channels, flexures and assemblies, flex rubbers and meshes, etc., please contact our sales team and we will be here to meet your needs!